viablebanner2

Space & Defense

Hi-Rel Commercial

Contact us

Home

Site Map

Back
Printable version
VTA02056-00
Main Unit Power Supply   
FEATURES          
-      Up to 600 Watts Output Power
-      Wide AC Input Range / EMI Filtered
-      IEC60601-1, IEC60950 Standards, compliant
-      Natural Convection Cooling
-      Powers a Remote DC/DC Multi-channel PSU and a Cooling/Heating System
-      Low Ripple & Noise @ 20 MHz B/W
-      E5CN Compact, Intelligent Digital Controller for Remote Temperature Control
-      Over Current / Short Circuit Protection
-      High Degree of Input/Output Isolation
-      Humidity Interlock Protection
-      FPD Over / Under Temperature Protection
-      AC Remote ON/OFF signal
This low voltage, custom-designed Power Supply Unit used in Medical Electrical Equipment applications features a low profile package of 12Ó L x 7Ó W x 5.5Ó H, or 1.3Watt/in3. It has excellent load regulation, input filtering, and low output ripple as well as current limiting.
 
Input Specifications: 
Steady State Input Voltage...................................... 200 Vrms ± 10% to 240 Vrms ± 10%
Voltage Surge.......................................................... Nominal voltage +15% for 120 sec.
Frequency............................................................... 50 / 60 Hz, 1¯
Inrush Current......................................................... 10 A peak maximum
Isolation (input/output to chassis)........................... Meet IEC 60950 Specification 
Electromagnetic Interference................................... Meet EN55011, classA, EN61000-4-, 2/2/3/4/6/11, ENV50204.
Remote ON Signal ................................................. 100 Vrms ± 10% to 120 Vrms ± 10%. Once Remote ON voltage is applied to J8,
                                                                                 Enable signal (Active Low) is generated at J9.
Humidity Interlock Protection Signal ..................... To maintain operation, contacts at J10 must Remain closed. Once J10 contacts
                                                                                open, J9 will be disabled causing Relay Unit to Power OFF.
                                                                                 Each Main Unit is shipped with mating connector on J10 normally closed. To use
                                                                                 Humidity Interlock feature, remove and discard mating connector and attach
                                                                                 Humidity Interlock Cable.     
FPD Over/Under Temperature Protection............... Main Unit provides through a digital temperature controller an over/under
                                                                                 temperature protection for FPD. As soon as FPD temperature is below or over 2
                                                                                 set temperature points, Main Unit shuts-down and latches off. To reset Main Unit,
                                                                                 allow FPD to return to normal temperature range and cycle Main Unit Power OFF
                                                                                 for seconds and Power ON.                                                                                 
Overcurrent Protection............................................Constant current limiting for output voltage ³ 19 Vdc ± 2 Vdc. PSU will latch off
                                                                                 when output voltage < 19 Vdc. To restore power, cycle input power OFF then ON.                                                                                
Output Specifications: 
CH1:      Powers the Relay Unit P.S. and Fans of the Cooling / Heating System.
               i.e., its output power is dependent on the input requirements of the Relay Unit, for a given load condition.
CH2:      Powers the Cooling / Heating System which maximum duty cycle is 75%. Its voltage polarity is (+) during cooling periods
and (-) during heating periods.
Nominal                Maximum                     DC Load               Ripple                 Maximum
Voltage                  Current                        Regulation             & Noise                   Power
 CH1: + 28VDC      10 A                           ±4%                < 250 mVp-p  
CH2: ± 28VDC      10 A pk (7.5A avg)       ±4%                < 250 mVp-p                
Over-voltage Protection.......................................... <120% of nominal level
Over-current Protection........................................... Constant current limiting down to 19 ± 2VDC, then it latch up.
Efficiency (full load)............................................... > 80% 
Physical Specifications:
Weight.................................................................... < 5 kgs
Enclosure................................................................ Aluminum
Finish...................................................................... Alodine Chemical film, Class 1, yellow
Mounting Position.................................................. Three positions. PSU to be mounted with fins pointing UP. If oriented in any other
                                                                                 position, ensure there ia a minimum pf Linear Feet per minute airflow.
Environmental Specifications:  
Air Temperature...................................................... Operating: +10¡C to +50¡C @ Full Load, Storage: -10¡C to +60¡C
 Atmospheric Pressure............................................. Operating: 70 to 106kPa, Storage: 50 to 106 kPa
Humidity................................................................. Operating: 30 to 85% non-condensing
                                                                                 Storage: 30 to 85% non-condensing 
Reliability: 
Reliability Prediction (MIL-HDBK-217F)..............   >70,000 hours              
Hi-Rel Commercial

VCD 02056-00 Rev. 01